Welch Allyn MicroTymp2 Tympanometer: VENDOR DESCRIPTION: Welch Allyn MicroTymp2 uses foreground and background contrast on its supertwist green LCD screen to make it easy to see at any angle. Word prompts direct the operator, facilitating testing. Asterisks alert the operator to gradient/width abnormalities. The MicroTymp2 autoerase feature means you don’t have to waste time erasing memory. Erasure is accomplished automatically as you store the next test.
Portable, hand-held unit detects otitis media and other middle ear disorders. Measures sound returned from eardrum under positive, negative, and atmospheric pressure to determine status of middle ear. Accurately identifies many middle ear disorders in asymptomatic cases. Results instantly displayed on LCD . 226 Hz frequency, 85 db HL intensity.
FEATURES Only one second for data—ideal for the uncooperative patient Readable at any angle with supertwist green LCD screen Word prompts guide operator during middle ear testing Data appears on screen as it happens Save time with automatic erase feature Comfortable—the lightweight, well-balanced, ergonomic handle is easy to hold Accurate diagnosis, referrals, monitoring, and easier consultations with parents. Results can be viewed in a tympanogram, numerical form, or interpretive messages. Instant LCD tympanogram allows the operator to view test results Lightweight and hand-held for complete portability Color-coded tips come in four sizes to fit any ear size
SPECIFICATIONS Probe Tone Frequency: 226Hz ±3% Amplitude: 85 ±3dB re 20μPa in an ANSI HA-1 (2.0cc) coupler Distortion: 5% maximum total harmonic distortion in an ANSI HA-1 (2.0cc) coupler
Pump Direction of sweep: positive to negative pressure Speed: 400 ±40daPa/s average during data acquisition period Pressure Measurement System Range: +200 to -400daPa Display resolution: 20daPa Accuracy: ±15% or ±10daPa, whichever is greater Compensation: Auto-zero every test cycle
Admittance Measurement System Range: 0.2 to 4.0mmho total 0.0 to 1.5mmho for Ya 0.2 to 2.5cc for ±200Vea
Weight MicroTymp 2 Handle: 0.61lb/0.28kg Battery (Welch Allyn #72900): 2.4 V nickel-cadmium rechargeable, 14-16 hour recharge time Full charge yields at least 300 double-ear tympanograms Probe Tip: Large MicroTymp Color-coded Probe Tip
RESOURCES Welch Allyn MicroTymp2 Brochure Welch Allyn MicroTymp2 User Manual