Fluke 80i-600a AC Current Clamp by ValueTronics International, Inc
- Item Condition:
- Brand:
- Model#:
- Location:
- Price:
Product Description
The Model 80i-600A is a clamp-on ac current probe designed to extend the current measuring capability of an ac current meter to 600 amperes.
A clamp-on, 1000-turn coil designed into the probe allows measurements to be made without breaking the circuit under test. The coil serves as the secondary of a current transformer.
The current carrying conductor being measured serves as the primary. Because of a high efficiency, quadrature type of winding used, wire size and the position of the wire within the probe jaws does not effect accuracy.
Input Current Range: 1A to 600A
Output: 1 milliamp per ampere of input current (1mA/A)
Accuracy: ±2% of reading, 50 Hz to 1 kHz ±3% of reading (typical), 30 Hz to 50 Hz or 1 kHz to 10 kHz.
Working Voltage: 750V ac rms maximum
Maximum Conductor Size: 50.8 mm (2 inch.)
Shunt (load) Resistance: less than 12 Ω at multimeter input to maintain specified accuracy
Typical Bandwidth: -10% at 10 Hz and 50 kHz (1A, 400 Hz reference, excludes multimeter response)
Usable Current Range: 0.1A to 2000A, 5 seconds maximum above 600A
Safety: Protection Class Class II as defined in IEC 348 and ANSI C39.5