  • OUTPUT module 30 V - 10 A - 60 W

    • Output Voltage range 0 to 30 V

    • Output Current range 0 to 10 A

    • Output Power 60 W

    • Overvoltage protection range 2 to 32 V


  • Accuracy (at calibration temperature)

    • Output Voltage ± 0.04 % of setting + 10 mV

    • Output Current ± 0.1 % of setting + 50 mA

    • Over Voltage Protection ± 0.5 % of setting + 150 mV


  • Temperature coefficient

    • Output Voltage (0.016 % of setting + 2 mV)/K

    • Output Current (0.05 % of setting + 2 mA)/K


  • Long term drift (in an 8 hour period)

    • Output Voltage 0.04 % of setting + 5 mV

    • Output Current 0.05 % of setting + 10 mA


  • Programming Resolution (12bit)

    • Output Voltage 7.5 mV/LSB

    • Output Current 2.5 mA/LSB

    • Over Voltage Protection 7.5 mV/LSB


  • Source Effect (for any line voltage change within rating)

    • Voltage ± (0.01 % of setting + 2 mV)

    • Current ± (0.02 % of setting + 2 mA)


  • Load Effect (load change 10 % to 90 % or 90 % to 10 % full load)

    • Constant Voltage ± (0.01 % of setting + 5 mV)

    • Constant Current ± (0.02 % of setting + 5 mA)


  • Load cross regulation Negligible for both voltage and current

  • Output Noise (PARD)

    • Constant Voltage pp/rms 15 mV/3 mV

    • Constant Current rms 10 mA

    • Switching Frequency approx. 20 kHz approx.


  • READBACK ACCURACY:(referred to the output terminals)

  • Accuracy (at calibration temperature)

    • Voltage ± (0.05 % + 25 mV)

    • Current ± (0.1 % + 30 mA)


  • Measuring Resolution (12 bit)

    • Voltage resolution to interface 7.5 mV

    • Current resolution to interface 2.5 mA

    • Voltage resolution to display 10 mV

    • Current resolution to display 10 mA

    • Max. Current at Reverse voltage 10 A

    • REMOTE SENSE > 0.25 Volt drop per lead (at open sense connections, the output voltage will not be more than 3 V above the programmed value)

    • DYNAMIC OPERATION ÆI=1 A between 1 and 10A between 0.5 and 5A between 1 and 10A

      • Settling band ± 50 mV

      • dI/dt < 1 A/μs

      • Overshoot < 150 mV

      • Recovery time < 1 ms


    • Programming response (measured with resistive load)

    • Up (0 V to max. output) 100 ms @ 2 A

    • Up (0 V to max. output) 50 ms @ (no load)

    • Down(max. output to 1 V) 100 ms @ 2 A

    • Down (max. output to 1 V) 50 ms @ (no load)


  • Reprogramming delay

    • Range 50 ms to 60 s

    • Resolution 1 ms

    • Default 50 ms

    • Step Interval time up to 60 s