Sorry, the search criteria can not be empty
Search

RTH Electronic Computer Controlled Creep Rupture School Laboratory Equipment by Jinan Precision Testing Equipment Co., Ltd.

  • Item Condition:
  • Brand:
  • Model#:
  • Location:
  • Price:
Jinan Precision Testing Equipment Co., Ltd.
86-531-87960408


Contact the seller Add to Inquiry Cart

Product Description

Quick Details

  • Place of Origin:Shandong, China (Mainland)

  • Brand Name:PTE

  • Model Number:RTH

  • Power:Electronic

  • Usage:Deformation Testing Machine

  • Max. Load:30kN, 50kN, 100kN, 200kN

  • Effective measuring range:1%-100%FS

  • Working Temp. range:400?--1200?

  • Length of effective soaking zone:150mm

  • Furnace pipe:90mm (Inner Dia.)*400mm (L)

  • Furnace:300mm (outer Dia,)*420mm (L);

  • Furnace life:?3000hours

  • Measure Control temperature meter:digital display, sensitivity: 0.1?;

  • Auto compensation error of cold end:?0.5?

  • Timing error:0.2%

Packaging & Delivery

Packaging Details:seaworthy packing and export wooden cases
Delivery Detail:within 20 working days after depoist

Specifications

Creep laboratory equipment
1.routine creep,stress-rupture test
2.stress relaxation test
3.cycle load test low cycling test

RTH  Computer Controlled Electronic Laboratory Equipment

Application:

RTH series is a kind of full electronic creep rupture Measuring Equipment to perform routine creep and stress-rupture test., and also can carry on stress relaxation test, high-temperature tensile test in short time, cycle load test (load spectrum test), low cycling test etc.

Specification

Test Load

Max. Load: 30kN, 50kN, 100kN, 200kN

Effective measuring range: 1%-100%FS;

Test load accuracy: ≤1%;

Resolution: 1/60000

High temperature furnace & Temperature-controller

1.Working Temp. range: 400°C~1200°C

2.Length of effective soaking zone: 150mm

3.Furnace pipe & electric furnace size:

4.Furnace pipe: φ90mm (Inner Dia.)×400mm (L),

5.Furnace: φ300mm (outer Dia,)×420mm (L);

6.Furnace life: ≥3000hours (Constant working time less than 1000°C)

7.Furnace Temperature accuracy (°C):

Test Temperature

Temperature fluctuation

Temperature gradient

≤600

±2

2

more than 600~900

±3

3

more than 900~1200

±4

4

8.Measure & Control temperature meter: digital display, sensitivity: 0.1°C;

9.Auto compensation error of cold end:≤0.5°C;

  To be matched freely with thermocouple.

  Moving speed of lower grip holder

10.Speed range: 0.005~50mm/min;

11.Error: ±0.5%

12.Timing error: ±0.2%


Item added to Inquiry Cart