The HP 85052A 3.5 mm calibration kit is designed to be used with the 8510 network analyzer system. It consists of open, short, and fixed load terminations, various adapters, a sliding load termination, and a data cartridge containing the calibration constants for the termination devices in the kit. Also included, unless optionally deleted, is a 3.5 mm connector gauge kit and a 5/16 inch torque wrench for use on the 3.5 mm connections.

  • Electrical
    • Fixed loads: >36 dB return loss, DC to 2 GHz
    • Sliding load: >42 dB return loss, 2 to 26.5 GHz (airline portion only)
    • 15 cm airline: >44 dB return loss, 2 to 26.5 GHz (airline portion only)
    • Short circuit: ±2.6 degrees, DC to 26.5 GHz, ±1° average deviation
    • Open circuit: ±2.6 degrees, DC to 26.5 GHz, ±1° average deviation
  • Options
    • Option 010: Add a 15 cm beadless airline (1250-1876), which is especially useful for calibrating time-domain applications
    • Option 020: Deletes the connector gauge kit and the torque wrench, which are essential but can be replaced by equivalent tools
Status Discontinued