  • Product Overview

The HP 4329A high resistance meter is a solid-state measuring instrument intended for many very high resistance test applications. The 4329A may be applied directly to the measurement of resistive elements such as high-valued resisitors, capacitors, and materials, or to the examination of volume and surface resistivities of solid or liquid materials using suitable resistivity cells. Covering a resistance range of 500 kΩ to 2 x 1016 Ω, the 4329A is provided with seven discrete test voltages for maximum flexibility over the resistance range. Automatic indication of meter and multiplier scales to be read is featured, providing rapid reading and straightforward operation.

The 4329A uses a "self-oscillating parameteric converter" circuit. A variable-capacitance diode bridge circuit serves as the parametric element.

The internal dc current measuring circuit, which is the heart of the instrument, is made available for current measurement. Use of the 4329A as a picoammeter is intended as a limited application because of the need to calculate errors. However, this added current measurement is a convenient complement to high-resistance testing.

The 4329A has a "recorder output" voltage that is proportional to meter deflection. This voltage varies from 0 to 100 mVdc for meter deflection from ∞ to full-scale.

  • Resistance range: 500 kΩ to 2 x 1016 Ω
  • Resistance accuracy: ±10% of reading except on Rmax range after initial ∞ and full-scale calibration. Accuracy depends on the deflection angle of the meter. At the low resistance end of each scale, typical accuracy is ~±3%, near center scale ~±5%. On all voltage ranges, if multiplier is set to Rmax, add 3% (Above effective upper limits, accuracy is not specified.)
  • Test voltages: 10 V, 25 V, 50 V, 100 V, 250 V, 500 V, 1000 V
  • Current range: 0.05 pA to 20 µA
  • Current accuracy: ±5% using ideal current source of infinite impedance. For finite source impedances, accuracy is determined by the range register, which sets meter input impedance. These vary from 104 Ω to 1011 Ω, depending on the range selected. This additional error can be large, and should be calculated if the current measurement capability of the 4239A is to be used
  • Recorder output: 0 to 100 mVdc, proportional to meter deflection, output impedance 1 kΩ 
Status Discontinued
Alternative 4339B High-Resistance Meter, DC